Instructions for Using the PT-83 Calibration Verification Reference (CVR)
Your PT-83 Pin Meter Calibration Verification Reference (sold separately) allows you to verify two calibration points to determine if your meter requires recalibration. We offer two CVR models: one designed for low-humidity conditions with 7% and 12% reference points, and another designed for high-humidity conditions with 12% and 22% reference points. This gives you the flexibility to choose the CVR that best aligns with your typical measurement needs.
Getting Started
- Power on the PT-83 by pressing and releasing the On/Hold button.
- Press the Mode button until Mode 2 is displayed on the screen.
- Remove the pin protector cap if it’s still attached.
Important Application Note
For each calibration point (7%, 12%, or 22%):
- One pin must make contact with the designated percentage contact point, and the other pin must make contact with the center circular contact on the CVR.
- Then switch pin positions and repeat the process (each pin must touch the percentage contact and center contact once).
- The reading should fall within the specified range during both tests.
How to Verify Each Calibration Point
7% Calibration Check
- Touch one pin to the 7% contact, and the other pin to the center contact.
- Reading should be between 6.9% and 7.1%.
- Switch pin positions and repeat to confirm.
12% Calibration Check
- Touch one pin to the 12% contact, and the other pin to the center contact.
- Reading should be between 11.9% and 12.1%.
- Switch pin positions and repeat to confirm.
22% Calibration Check
- Touch one pin to the 22% contact, and the other pin to the center contact.
- Reading should be between 21.9% and 22.1%.
- Switch pin positions and repeat to confirm.
Tips for Accurate Readings
- Ensure firm, direct contact between pins and contact points.
- Keep your fingers clear of the pins during testing.
- Avoid areas with high static, such as carpeted rooms or low-humidity environments.
Troubleshooting
If the reading is outside the specified range for any test:
- Verify that both pins are making proper contact with the correct contact points.
- Confirm the meter is in Mode 2.
- Repeat the test using the pin-switch method.
If any reading remains outside the expected range after confirming proper technique and mode setting, your PT-83 may require recalibration. Please contact us at info@bessemeter.com to send in the unit for service.